Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8: (1 year) CAL VER /5500 INSTRUMENT: Fluke 10: (1 year) CAL VER /5500 DATE: 22-Sep-98 AUTHOR: Fluke Corporation REVISION: $Revision: 1.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 109 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 10 Series Service Manual (PN 900824 March 1992, Rev. 1, 7/93). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 55_10_2w.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F V 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP No warm-up time requirement. 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 90%. 1.006 DISP Move switch to OFF position. 1.006 DISP 1.006 DISP Note: Erroneous result may occur if switch is not 1.006 DISP placed in OFF position before beginning tests. 1.007 HEAD {RESISTANCE TESTS} 1.008 DISP Move switch to the right position. 1.008 DISP Press SELECT for Ohms 1.009 PIC 55_10_2W 1.010 HEAD RESISTANCE TESTS: {400 Ohm Range} 1.011 5500 400 0.0Z +0.2U 2W 2.001 5500 400 1.0Z 0.2U 2W 3.001 5500 400 350.0Z 5.5U 2W 4.001 HEAD RESISTANCE TESTS: {4 kOhm Range} 4.002 5500 4 1.000kZ 0.016U 2W 5.001 HEAD RESISTANCE TESTS: {40 kOhm Range} 5.002 5500 40 35.00kZ 0.54U 2W 6.001 HEAD RESISTANCE TESTS: {400 kOhm Range} 6.002 5500 400 100.0kZ 1.6U 2W 7.001 HEAD RESISTANCE TESTS: {4 MOhm Range} 7.002 5500 4 2.900MZ 0.045U 2W 8.001 HEAD RESISTANCE TESTS: {40 MOhm Range} 8.002 5500 40 35.00MZ 0.70U 2W 9.001 HEAD {CONTINUITY TEST} 9.002 DISP Press SELECT for Continuity. 9.003 5500 2.000V 0.032U 2W 10.001 HEAD {AC VOLTAGE TESTS} 10.002 DISP Move switch to the center position. 10.002 DISP 10.002 DISP Press SELECT for VAC. 10.003 HEAD AC VOLTAGE TESTS: {4V Range} 10.004 5500 0.000V S 2W 10.005 MEMI Enter UUT reading in volts AC: 10.006 MEME 10.007 MEMC 4 V +0.003U 11.001 5500 4 0.010V 0.003U 50H SI 2W 12.001 5500 4 0.010V 0.003U 400H SI 2W 13.001 5500 4 3.500V 0.105U 50H SI 2W 14.001 5500 4 3.500V 0.105U 400H SI 2W 15.001 HEAD AC VOLTAGE TESTS: {40V Range} 15.002 5500 40 35.00V 1.05U 50H SI 2W 16.001 5500 40 35.00V 1.05U 400H SI 2W 17.001 HEAD AC VOLTAGE TESTS: {400V Range} 17.002 5500 400 350.0V 10.5U 80H SI 2W 18.001 HEAD AC VOLTAGE TESTS: {600V Range} 18.002 5500 600 600V 20U 400H SI 2W 19.001 HEAD {DC VOLTAGE TESTS} 19.002 DISP Press SELECT for VDC. 19.003 HEAD DC VOLTAGE TESTS: {4V Range} 19.004 DISP Press RANGE to select the 4V Range. 19.005 5500 4 0.000V 0.002U 2W 20.001 HEAD DC VOLTAGE TESTS: {40V Range} 20.002 DISP Press RANGE to select the 40V Range. 20.003 5500 40 0.02V 0.01U 2W 21.001 HEAD DC VOLTAGE TESTS: {400V Range} 21.002 DISP Press RANGE to select the 400V Range. 21.003 5500 400 -0.2V 0.1U 2W 22.001 HEAD DC VOLTAGE TESTS: {4V Range} 22.002 DISP Press RANGE for 2s to select autorange. 22.003 5500 4 3.500V 0.055U 2W 23.001 HEAD DC VOLTAGE TESTS: {40V Range} 23.002 5500 40 -35.00V 0.54U 2W 24.001 HEAD DC VOLTAGE TESTS: {400V Range} 24.002 5500 400 100.0V 1.6U 2W 25.001 HEAD DC VOLTAGE TESTS: {600V Range} 25.002 5500 600 -600V 10U 2W 26.001 END